Figure 24Set Alarms
5K-MNL-GEM5000 GEM5000 Gas Analyzer
Select the soft-key ‘Filter’ to filter the data by sample point ID, or specify before or after date. Press the
soft-key ‘Exit’ to exit the filter menu and return to the ‘View Data’ screen. The screen shown in Figure 23
will filter for the ID named “ANAW0001”. All dates in the instruments history will be used because an
“After Date” and a “Before Date” is not specified 00/00/00.
Figure 23Filter Data
4) Select the soft-key ‘Delete’ followed by the appropriate soft-key to delete a single
reading or all filtered readings. Press soft-key ‘Cancel’ to cancel the deletion request.
5) Select the soft-key ‘Exit’ to exit the view data screen.
7.4.4 Set alarms
This option enables the operator to define the conditions for which an
alarm will be triggered. These conditions apply to the general operation of
the instrument and are not ID specific. A summary of the alarm settings
can be found in ‘Key 3 – Summary’.
1) Select the ‘Menu’ button on the front of the analyzer to display the ‘Device Settings’
menu. Press the soft key to display ‘Operation Settings’.
2) Select ‘Key 4 – Set Alarms’ and the following menu is displayed:
Page 74 Copyright LANDTEC North America, Inc. V1.13