5K-MNL-GEM5000 GEM5000 Gas Analyzer
13 Service
The GEM5000 gas analyzer should be regularly serviced to ensure correct and accurate
operation. LANDTEC recommends a service and recalibration every 12 months.
The GEM5000 analyzer is CSA and ATEX certified for use in potentially explosive areas. As
such it should be serviced only by qualified engineers. Failure to observe this will result in
the warranty becoming invalid and could invalidate the CSA, ATEX, and other certifications.
Warning
If the GEM5000 is to be serviced by trained LANDTEC personnel.
Service by any untrained personnel will negate service
technicians serviced by unqualified engineers the CSA, ATEX, and
other certifications may be invalidated and the instrument may be
unsafe for use in a potentially explosive atmosphere.
User serviceable parts:
There are no user serviceable parts inside the instrument.
The following parts can be user serviced:
In-line water filter
This should be regularly inspected for obstructions, moisture
or damage and changed if needed. The instrument should
never be operated without the in-line water filter as this
may result in water entering the instrument.
Sample tubing
Always ensure that sample tubes are not contaminated
or damaged.
Gas port connectors
Periodically check that the O-rings on the gas port
connectors of the hoses are not damaged. A damaged O-
ring can let air into the sample gas and result in incorrect
readings. If the O-ring it should be replaced immediately.
H2S filter material
When the filter material changes color to a light grey
color the filter should be replaced.
Page 114 Copyright LANDTEC North America, Inc. V1.13