5K-MNL-GEM5000 GEM5000 Gas Analyzer
Once the calibration is complete, press Exit to return to the gas check screen.
Note: A ‘Simple Gas’ reading is a way of creating a recordable and downloadable
composition only reading to document a calibration check event. Refer to 8.3 Special action
for more details about ‘Simple Gas’ Readings.
9.6.1 Adding / Deleting Gas Mixtures
If your instrument is capable of measuring gases other than Methane, Carbon Dioxide, and Oxygen, then
it will be necessary to define the additional span gases that you will use to calibrate your instrument. The
following will demonstrate how to add a gas mix and delete a gas mix.
Adding a “Span Target” for H2S
From the “Select Gas Mix” screen choose ‘Add’
The Gas Mix Screen will appear. Press the key to allow for the editing of the Mix Name. Typically the
last mix name defined will appear. Press any key to change the mix name and the Delete Key will
appear. You can also press the center for ‘Delete’.
Page 104 Copyright LANDTEC North America, Inc. V1.13