April 8, 2011 HN FAQ Memo, Version 3.0 Page 11 of 108
1. Incorrect mounting of Bd1, or Bd1 has been exposed to water.
2. Coatings in main line tube (ø2mm) causing malfunction of Bd5, could be caused by cleaning problems.
Ready For CIP Timeout:
1. Error in heating the CIP containers in the SI
Ready For CIP Timeout+Fill Timeout:
1. Error in water supply or water level sensors in CIP containers in SI
Fill Timeout:
1. Error in water supply or water level sensors in CIP containers in SI
No AI-CIP water:
1. Error in water supply or water level sensors in CIP containers in SI
Bd5 eller Bd1 CIP Air Detect:
1. Air has been detected in Bd5 or Bd1 during cleaning,
2. Check HSM->Climate->cleaning diagram below. to see if it is a bubble or missing water.
3. If it is a bubble it is OK
4. if missing water ict could be an error in water supply or water level sensors in CIP containers in SI
SI Detergent Empty:
1. Empty Detergent container, or an air bubble has been detected by the detergent Bd.