Step.3
Disconnect the tube from the SI to the AI at the connection app. 1m
from the AI. Place a bucket or similar under the end of the tube to
collect the water leaving the tube during cleaning.
Request a Cleaning from the SI, (SI menu 4.2.3.2.6 Req. CIP).
Check tube for coatings and repeat cleaning if necessary.
Reconnect the tube with the tube to the AI, and request a new
Cleaning. (SI menu 4.2.3.2.6 Req. CIP).
Check that water flows both from the dosage needle and from the
mainline out tube.