April 8, 2011 HN FAQ Memo, Version 2.0 Page 37 of 108
Step.3
Disconnect the tube from the SI to the AI at the connection app. 1m
from the AI. Place a bucket or similar under the end of the tube to
collect the water leaving the tube during cleaning.
Request a Cleaning from the SI, (SI menu 4.2.3.2.6 Req. CIP).
Check tube for coatings and repeat cleaning if necessary.
Reconnect the tube with the tube to the AI, and request a new
Cleaning. (SI menu 4.2.3.2.6 Req. CIP).
Check that water flows both from the dosage needle and from the
mainline out tube.
Check tube for
water flow from
needle.
HN1.
Check tube for
water flow from
needle. HN2.