April 8, 2011 HN FAQ Memo, Version 2.0 Page 41 of 108
Clean the channels in the flow module manifold with a pipe cleaner and a
detergent solution. Inspect and ensure that all impurities and coatings are
removed.
Reassemble module, and remount on the AI.
Step.6
Replace the tubes from the flow module to the dosage head if there is any signs
of coatings or bends ( AI dosage head tubes)
Ø1 tubekit
60005192
Clean buffer valve seat.
Clean all channels, take care
of 0-rings.