7.6.5 Submenu LMDkt Æ O0c-- (Master): Setting internal control parameters
You only obtain the following possible settings when the control variable is set to "Internal" (⇒ Section
7.7.4. See also 7.6.6.)
Set control parameters
O0c--
°C
Display flashes. Set
with or .
Xp range: 0.3K … 25.0 K.
Proportional range [K]
^O &/
°C
* Important:
The windows for setting the
internal control:
rs (=tv) and rc (= td)
are only shown when rsc
(=tvd) has been changed
from the works setting
"Automatic" to "Manual".
Changeover should only be
carried out by experienced
control technicians.
Display flashes. Set
with or .
Tn range: 3s … 180s, Off.
Reset time [s]
rl 2/
°C
*Display flashes. Set
with or .
Tv range: Off, 1s … 145s.
Deriv. time [s]
rs 10
°C
*Display flashes. Set
with or .
Td range: 0.0s … 25s.
Damping time [s]
rc §4
°C
longer
than
Display flashes. Change tv
and td from
@=Autom. to
LM=Manual
with or .
Auto/manual for tv, td
rsc @
°C
The works setting (default)
of the internal control
parameters has been
restored and
cmlD is
displayed.
longer
than
Control works settings
cDE
°C
End
Dlc
°C