Proline Low-temperature Thermostats
60 Starting up YACE0072 / 21.08.07
− Select the desired resolution with or
.
− Accept selection with or End ,
or quit the window with without making
changes.
7.8 Special settings
7.8.1 Setpoint resolution
This function enables the resolution of the setpoint T
set
to be increased from the standard value of
0.1 °C to 0.01 °C.
Master
R
3x to LMDls
to
O@p@ 1x
− Call setpoint resolution
R .
− The current setting is displayed (here
<0 for the works setting of
0.1°C).
− The resolution display
<0 flashes.
or
−
<0 for 0.1°C
−
</0 for 0.01°C
Wait 4 seconds or
− Display flashes 4 s Æ new value is automatically accepted, or
− the value is immediately accepted with the Enter key.
7.8.2 Defining the type of start mode
Usually it is desirable that the thermostat carries on operating again after an interruption in the voltage
supply. However, if for safety reasons you do not wish this, you can insert an intervening manual
activation step.