26
Setting up the options
Positioning the filter block slide*
• Hold the filter block slide so that the warning
label is positioned is located at the front left
side and insert it into the dovetail holder (26.2)
on the left side of the microscope stand.
• The filter block slide can now be moved
between the three switching positions.
Attention!
The filter block slide is not protected against
inadvertent sliding out.
Attention!
When using an incident-light dichroic mirror
or polarisation dichroic mirror in conjunction
with fluorescence filter blocks, there is a risk
of glare in case of inadventent switching
between positions!
!
Note
These assembly steps are not required if no ad-
ditional accessories have been purchased with
the microscope.
Assembling the fluorescence filter blocks*
Note
For microscopes with integrated incident-light
fluorescence facility only.
The filter block slide (Fig. 26) accommodates up
to three fluorescence filter blocks.
• For assembly of the filter blocks, remove the
slide cover (26.5).
• Insert the filter blocks (26.1) into the dovetail
holder (26.2), with the engraved lettering
pointing upward. Make sure that the filter
blocks click in place.
• Reinstall the slide cover.
• Check that the slide cover (26.5) is properly
installed.
• Mark the filter positions using the enclosed
adhesive labels.