• Place a new 12 V/100 W halogen lamp straight
into the lamp holder (34.1).
• Move the collector back to its original
position.
• Replace the lid and fix it with the screw (31.1,
32.1).
• Connect the lamp housing to the power unit:
Assembling the 106 z* lamp housing with
halogen lamp
Note
For microscope with integrated incident-light
fluorescence facility only.
Fig. 34 LH 106 lamp housing, opened
1 Lamp holder with 12 V/100 W halogen lamp, 2 Collector,
3 Diffuser
Fig. 35 107/2 lamp housing, opened
1 Collector, 2 Lamp holder with 12 V/100 W halogen lamp
21
Attention!
Prior to making any assembly work, always
disconnect the power supply at the external
transformer outlet and the microscope
outlet!
• Unscrew the lock screws (36.4 and 36.9) with
a crosstip screwdriver.
• Slightly pull out the cut-out plug (36.11) from
the socket and flip up the lid (36.1).
Attention!
Leave the protective covering on the lamp
until the lamp is in its holder! Avoid making
finger prints or wipe them off immediately.