EasyManua.ls Logo

Leica DMIL - Positioning the Transmitted-Light Illumination Carrier

Leica DMIL
244 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
18
Positioning the transmitted-light
illumination carrier
• Position the transmitted-light illumination car-
rier into the column from above while
pressing on the notched lever for the con-
denser level adjustment (9.5).
• Position the transmitted-light illumination
carrier (9.3) on the transmitted-light illumination
column (9.4) depending on the condensor used
(S 55 or S 90) and release the notched lever.
The markings (9.6) are referred to a liquid
level of 15 mm. For microscopes with dual
markings, the lower line indicates a liquid
level of 15 mm and the upper line indicates a
liquid level of 50 mm.
• Make sure that the transmitted-light illumi-
nation carrier is locked in place.
Fig. 9 Transmitted-light illumination unit with condenser
1 Condenser (S 90), 2 Condenser holder, 3 Transmitted-light
illumination carrier, 4 Transmitted-light illumination column,
5 Notched lever for condenser level adjustment, 6 Markings
Note
A screw (6.1) on the transmitted-light illumi-
nation column prevents that the condensor
collides with the specimen stage.
Repositioning the transmitted-light illumination
carrier with the specimen stage
Note
For the heating stage* only one position of the
transmitted-light illumination carrier is possible.
This position is determined by the existing holes
in the stage.
Repositioning of the transmitted-light illumina-
tion carrier is only possible with the DM ILB or
DMILT tubes.
The specimen stage can be made accessible
from three sides by repositioning it by 180°.
1
3
4
2
5
6
6

Table of Contents