70
Instructions for Use, V 2.2 RevI – 09/2016
7. Troubleshooting
Problem Possible cause Corrective action
The blade is dull.
The pressure plate is damaged
or incorrectly adjusted.
Clearance angle of the knife/
blade too small.
Laterally displace the knife
holder or insert a new blade.
Insert a new pressure plate or
use a new knife holder.
Methodically experiment with
larger clearance angle settings
until you have found the optimum
angle.
2. Section compression
The sections are very com-
pressed, show folds or are
squeezed together.
The blade is dull.
The specimen is too warm.
The sectioning speed is too fast.
Use another area of the blade or
a new blade.
Cool the specimen before sec-
tioning.
Reduce the sectioning speed.
3. "Stripes" in sections
For knife holder E
There is an accumulation of par-
affin on the rear pressure plate
of the knife holder.
Remove paraffin from this area
on a regular basis.
4. Noises during sectioning
The knife "sings" when section-
ing hard specimens. The sec-
tions show scratches or chatter
marks.
The sectioning speed is too fast.
The clearance angle is too
wide.
Insufficient clamping of speci-
men clamp and/or knife holder.
Select a slower sectioning
speed.
Methodically reduce the clear-
ance angle setting until you have
found the optimum angle.
Check all screw and clamp con-
nections on the specimen holder
system and the knife holder. If
necessary, tighten the levers and
screws.
5. High blade consumption Too great of a sectioning force
was applied.
Adjust the sectioning speed and/
or section thickness when sec-
tioning. Select a thinner section.