EasyManua.ls Logo

Leica RM2255 - Possible faults

Leica RM2255
82 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
69
Leica RM2255
7. Troubleshooting
7.2 Possible faults
Problem Possible cause Corrective action
1. Thick/thin sections
The sections alternate between
being thick and thin, or there is
chattering in the sections, or the
specimen is torn out of the em-
bedding. In extreme cases, there
are no sections whatsoever.
Blade, knife holder or orientation
is not clamped properly.
The specimen is not clamped
properly.
Reclamp the blade, knife holder
or orientation.
Check that the cassette is se-
curely clamped in the universal
cassette clamp.
If the universal cassette clamp
is soiled by wax, clean the UCC
(refer to Chapter 8.1 - "Cleaning
and Maintenance - Universal
cassette clamp").
When using cassettes whose
lid is molded on, check that the
broken edge allows the cassette
to be clamped securely; if neces-
sary, remove burrs or clamp the
cassette into the universal cas-
sette clamp horizontally instead
of vertically.
If the cassette dimensions are
within the specified
tolerances and the cassette still
cannot be securely clamped into
place, it may be that the univer-
sal cassette clamp is configured
incorrectly or is defective. In this
case, have Technical Service
inspect and reconfigure the
universal cassette clamp.
When using cassettes, par-
ticularly thin-walled cassettes,
from a company other than
Leica or Surgipath, the cassette
may become deformed or other
clamping problems may arise.
If while trying to clamp the cas-
sette you realize that it is not
securely clamped into place,
an alternative tensioning clamp
must be used.

Table of Contents

Other manuals for Leica RM2255

Related product manuals