2.6 Maximum Static Pressure for Pump Heads
Figure 8: Specification for maximum static pressure of pump heads
(Pressure limits only valid for pump head mounted on motor.)
0
20
40
60
80
100
120
140
77 87 97 107 117 127 137 147 157 167 177 187
0
1
2
3
4
5
6
7
8
9
10
25 30 35 40 45 50 55 60 65 70 75 80 85 90
[psi]
F]
[bar]
Liquid Temperature in [°C]
Max. Static Pressure for LPP-600.7/18 (PVDF)
Max Static Pressure for LPP-600.27 (Stainless Steel)

Limitation: Given by deformation. Burst pressure is significantly higher
Ambient Temp = 25 C
Note: Pressurelimit only valid for pump head mounted on motor