22
LT360 Precision Turntable
DLL Programming Manual
■ Class Set (out)
■ Identifier lt_SetBaudRate
■ Hex Value 0x21
■ Call LT360LIB_CmdValue(Handle, lt_SetBaudRate, 'nnnnn');
■ Parameter Baud rate, eg. '9600'
Description
This command sets the baud rate for RS-232 communications. Once
the new baud rate is set, all further RS-232 communication with an
LT360 must be at the new baud rate.
Legal baud rate values are:
9600
14400
19200
38400
57600
Note: Increasing the baud rate will not necessarily produce a direct effect
on the speed of communication and commands with the LT360. The
LT360 has a response time for commands which is dependent on many
factors internally. The CPU must perform many other tasks so there can
be a 1-25 mSec latency before the commands are processed regardless of
the baud rate.
2.6 Set Baud Rate