LT360 Precision Turntable
DLL Programming Manual
Class Get (in)
Identifier lt_GetDueDate
Hex Value 0x99
Call LT360LIB_CmdValue(Handle, lt_GetDueDate, RtnStr);
Returns 'mmm-dd-yyyy'
Description
This command returns the date when the LT360 should be re-calibrated.
For example: JAN-01-2007 The date value is returned as a string.
Calibration is performed using the Win32 application software. Calibra-
tion is performed originally at the factory, and in most cases will be
permanent and not be required again. However if your application
requires NIST compliance, this is the date when the LT360 should be
rechecked. It is possible for a customer or other lab to performed the
calibration, but only if the proper equipment is available. The computer
must have a GPIB bus, and two 6-1/2 digit DMMs are required them-
selves with accurate recent calibration. See the LT360 Manual for
further details.
2.29 Get Calibration Due