Loading...
Project SERVICE MANUAL
Date
05-2017
Version 05
78/84
Battery
Battery
Test the battery functions.
Test the internal resisitance of the battery.
Replace the battery after two years.
N/A
Electrical
Electrical test
Electrically test the bed according to the appropriate standards.
Write the results to the preventative check report.
Electrical
Set up the next preventive check
Set up the next preventive check on the LCD screen.
N/A
Label
Placing the preventative check sticker on the bed
Place the LINET sticker on the foot end crossbar and make a note of
the date of the next preventative check (1 year).
Report
Complete the preventative check report
Preventative check
report