11.2.5 APT Mode
time. APT Mode is intended to be used as an aid to patient care in prevention and treatment of skin breakdown related to
pressure damage.
To select APT mode:
Press button 13 once or repeatedly until indicator 4 comes on.
WhenAPTmodeisswitchedon:
Indicator 4 lights white.
Pressure indicator 3
Once the pre-set pressure is reached, indicator 3 will remain solid on.
NOTE After 3 cycles of APT Mode with deated individual cell an alert indicating fail to inate is triggered.
11.2.6 MaxMode
and the mattress replacement system will switch back to APT or CLP mode automatically. Selected Max Mode interrupts
the return to the previous mode.
ToselectMaxMode:
Press button 13 once or repeatedly until indicator 5
IfMaxModeisrequiredforlessthan30minutes:
Set new mode by pressing button 13.
IfMaxModeisrequiredformorethan30minutes:
Press MODE Button once or repeatedly until Max Mode is set again after 25 minutes of Max Mode.
WhenMaxModeisselected:
Indicator 5
Once the set pressure is reached, indicator 5 will remain solid on.
During the last 5 min, an alert signal will sound every minute.
WhenMaxModeisselectedforthesecondtime:
During the last 5 min, an audible reminder will sound every minute to remind the personnel that Max Mode
will soon end automatically.
NOTE It is not possible to select Max Mode a third time. Selecting Max Mode again is only possible when APT Mode or
CLP Mode has been running for at least 30 min.
11.2.7 CLPMode(ConstantLowPressureMode)
CLP Mode (constant low pressure mode) keeps the mattress pressure at the selected level. The pressure is checked
every 30 seconds, and adjusted if necessary. CLP Mode is intended to be used as an aid to patient care in prevention of
skin breakdown related to pressure damage.
To select CLP Mode:
Press button 13 once or repeatedly until indicator 14
WhenCLPModeisswitchedon:
CLP indicator 14 will light solid and indicator 3
Once the pre-set pressure is reached, indicator 14 and indicator 3 will remain solid on.