Description Default Profile Range/Option Your requirement
Standard Alarm Parameters
Upper Alarm
Alarm direction upper upper Upper
Trigger this alarm disabled enabled or disabled
Temperature threshold value
3
- -25°C to 6C (-13°F to 140°F)
Alarm activation type
4
instant instant, accumulative or consecutive
Activation delay time - 1 minute to 45 days
Lower Alarm
Alarm direction Lower Lower Lower
Trigger this alarm disabled enabled or disabled
Temperature threshold value - -25°C to 6C (-13°F to 140°F)
Alarm activation type instant instant, accumulative or consecutive
Activation delay time - 1 minute to 45 days
Multi-AlarmParameters
5
Alarm 3
Alarm direction - upper or lower
Trigger this alarm disabled enabled or disabled
Temperature threshold value - -25°C to 6C (-13°F to 140°F)
Alarm activation type instant instant, accumulative or consecutive
Activation delay time - 1 minute to 45 days
Alarm 4
Alarm direction - upper or lower
Trigger this alarm disabled enabled or disabled
Temperature threshold value - -25°C to 6C (-13°F to 140°F)
Alarm activation type instant instant, accumulative or consecutive
Activation delay time - 1 minute to 45 days
Alarm 5
Alarm direction - upper or lower
Trigger this alarm disabled enabled or disabled
Temperature threshold value - -25°C to 6C (-13°F to 140°F)
Alarm activation type instant instant, accumulative or consecutive
Activation delay time - 1 minute to 45 days
Alarm 6
Alarm direction - upper or lower
Trigger this alarm disabled enabled or disabled
Temperature threshold value - -25°C to 6C (-13°F to 140°F)
Alarm activation type instant instant, accumulative or consecutive
Activation delay time - 1 minute to 45 days
Advanced Display Options
Alarm remains active when readings return to spec enabled enabled or disabled
Clear alarm with START/Mark button
enabled enabled or disabled
Pause alarm/statistics processing for X readings 0 (disabled) 0 (disabled) to 255
Switch off display after 30 seconds (Power save) enabled enabled or disabled
Show alarm screens during review disabled enabled or disabled
Allow reset of the recorder with the start button disabled enabled or disabled
3
Upper value must be above lower value
4
l Instant = one temperature reading is above (below) the threshold
l Consecutive = temperature readings are above (below) the threshold for the time defined in the activation delay without interruption
l Accumulative = temperature readings are above (below) the threshold for the total time defined in the activation delay time, but may not necessarily be sequential.
5
Only on model "M" with multiple alarms.
Page 19 of 21 UTRID-16 Product User Guide, Revision 1, copyright © 2004-2016, LogTag
®