UTRID-16 Start Options
During configuration with LogTag
®
Analyzer you can decide when
the UTRID-16 starts taking temperature readings.
l ... via Push button start: The recorder will start taking
temperature readings after you have gone through a specific start
sequence of pressing and releasing the START/Mark button (see
Starting the Recorder on page1).
l ... via Date/Time start : The recorder will start taking
temperature readings at the date and time you specify (local
time).
When you choose the push button start option, you can select to
record pre-start readings or Begin recording after a delay.
l If you select pre-start readings, the UTRID-16 starts recording as
soon as it is unplugged from the USB port
4
and will continue to do
so until you start the unit. No alarms are processed while pre-
start readings are being taken, pre-start readings will not appear
in the PDF file, and no PDF file will be generated. Using pre-start
readings is a good way do avoid data loss if you forget to start the
unit, as you can stillaccess the data using LogTag
®
Analyzer.
l If you configure the UTRID-16 to start after a delay period, the
recorder will not immediately record temperature readings after
you have pressed START/Mark, but start a countdown timer
instead, and record readings only after the timer has ended. It
will, however, continue to take pre-start readings, if enabled.
You cannot combine a date/time start with pre-start readings or the
start delay function.
4
The first pre-start reading is taken one log interval after you unplug the unit from
the USB port.
Page 7 of 21 UTRID-16 Product User Guide, Revision 1, copyright © 2004-2016, LogTag
®