LeadCare
®
II User's Guide 1-11
Intended Use
The LeadCare II Blood Lead Analyzer and Test Kit provide a measurement of the
amount of lead in a fresh whole blood sample. The LeadCare II Blood Lead
Analyzer is intended for in vitro (external) use only. It is for lead testing only. The
test kit components are designed for use only with the LeadCare II Blood Lead
Analyzer. This test system is for professional use only.
Operating Requirements
CAUTION: Do NOT place the LeadCare II Blood Lead
Analyzer in a drafty area. For example, do NOT place the
analyzer near air conditioning or heating vents. If the
temperature is out of operating range, or if the temperature is
unstable, the following messages appear on the display.
TEMP IS TOO HOT
PLEASE WAIT UNTIL
ANALYZER IS IN
TEMP RANGE
TEMP IS TOO COLD
PLEASE WAIT UNTIL
ANALYZER IS IN
TEMP RANGE
If the temperature is unstable, the WARNING message appears on the display
and flashes on for 2 seconds. Move the analyzer to a more suitable location and
try again later.
WARNING
TEMP IS UNSTABLE
TEST MAY FAIL