II User's Guide 2-3
Analyzer Self-Tests
When you first turn on the analyzer, you will hear a beep and see the startup and
self-test messages.
The analyzer performs a series of self-tests. The LeadCare II self-test is a set of
internal electrical and software checks that ensure the proper operation of the
system's electronic components. The purpose of the tests is to ensure that each
critical hardware component of the system is operating at the correct level. If any
one component of the system fails this initial self-test, the user is warned that the
unit requires service and the user is prevented from running any patient samples.
The first time you turn on the analyzer, the screen reads:
PLEASE CALIBRATE
ANALYZER WITH BUTTON
Once the analyzer has been calibrated, the following message appears.
CALIBRATION
SUCCESSFUL
IF NEW LOT
TEST CONTROLS
PREPARE SAMPLE
USE SENSOR LOT XXXX
OR RECALIBRATE
THEN INSERT SENSOR
You can also turn on the analyzer by inserting a sensor.
If you insert a sensor to turn on the analyzer, the following message appears:
ADD SAMPLE
TO X ON SENSOR
SENSOR LOT XXXX