Magellan LeadCare II - Control Test Results Above the Target Range: Possible Causes

Magellan LeadCare II
70 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
LeadCare
®
II User's Guide 5-5
Testing Conditions
Do NOT operate the LeadCare II System in drafty locations.
Make sure that the analyzer and the test kit are maintained at a constant
temperature. If you move any part of the system from one place to another, (for
example, from outside into a laboratory) wait for the analyzer and components to
reach a stable temperature.
User Technique
Do NOT touch the black bars on the sensor.
Make sure the sensor is inserted under the sensor guides and sits flush on the
sensor deck.
Do NOT touch the sensor while the analyzer is running a test.
Make sure the control and treatment reagent is thoroughly mixed before placing
onto the sensor. The mixture will appear red.
Control Test Results Above the Target Range: Possible Causes
Possible causes of high control test results include:
Control lot number does not match the sensor lot and calibration code.
More than 50 µL of control material was transferred to treatment reagent.
The test sample is warmer than detected by the analyzer.
The analyzer is not properly calibrated.
See detailed precautions below:
Control Sample Preparation
Use the capillary tube and plunger to transfer 50 µL of blood into the treatment
reagent vial.
Wipe excess sample off the outside of the capillary tube with a clean gauze pad or
laboratory wipe before adding the control to treatment reagent.
Equipment Setup, Calibration and Testing Materials
Make sure the controls were properly stored: Room temperature storage is
considered 15 - 27ºC (59 - 80ºF).
Check the expiration date on the control vial to verify the controls have not expired.
Make sure that the lot number of the control matches the lot number on the sensor
container and the lot number on the display screen.
NOTE:
The controls are only intended for use with the test kit in which
they come. When the other reagents included in the test kit are
used up, discard the controls.