LeadCare
®
II User's Guide 5-6
Make sure the analyzer is calibrated properly. Use the calibration button supplied
with the test kit. When processing the sample, check to make sure the code on the
analyzer screen matches the code of the calibration button for the test kit.
Testing Conditions
Do NOT operate the LeadCare II System in drafty locations.
Make sure that the analyzer and the test kit are maintained at a constant
temperature. If you move any part of the system from one place to another (for
example, from outside into a laboratory) wait for the analyzer and components to
reach a stable temperature.
User Technique
Do NOT leave the treatment reagent tube uncapped other than to add the control
sample or to remove the sample for testing.
Do NOT touch the black bars on the sensor.
Do NOT touch the sensor while the analyzer is running a test.
Make sure the control and treatment reagent is thoroughly mixed before placing
onto the sensor. The mixture will appear red.
Screen Display Messages
There are a number of standard screen display messages that appear during the
routine testing procedure. However, other messages may appear if the analyzer
detects a condition that prevents normal operation. The following table shows
some of the display messages.
NOTE:
Occasionally, error messages not noted in this guide may
appear on the display. Please note what error message was
displayed and call Product Support for additional instructions.
New users may want to check that the sensor was completely
inserted into the analyzer first. Product Support can be
reached at 1-800-275-0102.