II User's Guide 5-8
Maintaining the Analyzer
Cleaning the Analyzer
Remove used sensors from the analyzer as soon as a result is recorded.
Clean the analyzer with a damp cloth and warm, soapy water. Do NOT
immerse in water.
Disinfect with dilute (10%) bleach solution.
Do NOT leave any soap film on the analyzer.
Do NOT allow liquid of any kind into the sensor connector.
Do NOT get the metal pins in the sensor connector wet.
Do NOT wash the inside of the calibration button reader.