LeadCare
®
II User's Guide 4-3
WARNING:
Blood can transmit infectious diseases. Use universal
precautions while collecting and handling blood samples.
Follow the procedures set up by your institution for
meeting federal, state, and local regulations.
Dispose of sensors, capillary tubes, plungers, and
droppers in biohazard container.
Use caution when handling the LeadCare II Treatment
Reagent. This reagent contains dilute hydrochloric acid.
Refer to the LeadCare Treatment Reagent Safety Data
Sheet that appears in Appendix E of this manual for safe
handling instructions.
CAUTION:
Do NOT use sensors that have been dropped, previously
handled, broken, scratched or damaged in any way. This
could cause inaccurate test results.
Make sure the sensor is inserted under the sensor guides
and sits flush on the sensor deck. Inserting the sensor
above the guides could cause inaccurate test results.
Do NOT use any test kit or its components past the
expiration date. This could cause inaccurate test results.
Do NOT leave the treatment reagent vial uncapped other
than to add the sample or remove the sample/reagent
solution. The tube and its contents could be
contaminated causing inaccurate test results.