04/04/2019
To ensure compliance with the EMC requirements as specified in IEC 60601-1-2, it is
essential to use only the following accessories (see Table 22).
Table 22 EMC Requirements Accessories
ITEM
MANUFACTURER
MODEL
ABR Preamplifier
MAICO
-
OWA Probe
RadioEar
-
RadioEar
RadioEar
IP30
IP30 50 Ohm stereo ID headset
RadioEar
IP30
BERAphone
®
MAICO
easySceen
Conformance to the EMC requirements as specified in IEC 60601-1-2 is ensured if the
cable types and cable lengths are as specified in Table 23.
Table 23
EUT SUPPORT EQUIPMENT
Item
Manufacturer
Model
Cable
SIP/SOP
Serial no.
Length
[m]
Screened
[Y/N]
Socket ID
Type
Power
Supply
UE / Fuhua
UE08WCP-
050160SPA
-
-
Mains power
AC
supply
-
1,5
N
Micro USB on
the wireless
charger/cradle
DC
supply
Wireless
charger/
cradle
MAICO
-
-
-
-
-
-
Audiom
etric
Insert-
Headset
(50Ω)
Radioear
IP30
0,25
Y
On the
preamp:
Socket
marked with
ear symbol
Analog
output
Serial
data
ID028384
Ear
Probe
RadioEAR
OWA
0,48
Partial
On the preamp:
Socket
marked with
ear symbol
or top socket
on the Eris
device
Analog
output
Mic input
Serial
data
-
Preamp
MAICO
-
1,15
Partial
Top socket on
the Eris device
Analog
output
Mic input
Serial
data
MA9017639
Electrode
cables
Sanibel
Supply
-
0,51
N
On the
preamp:
Colour marked
sockets with
head symbol
Analog
input
for
Physio-
logical
signals
-