Operation Manual easyScreen
The following table shows the protocols that can be installed in HearSIM™
ABRIS A00 CE-Chirp 35 dB nHL
CE-Chirp
®
stimulus of 35 dB nHL (default)
ABRIS A01 CE-Chirp 30 dB nHL
CE-Chirp
®
stimulus of 30 dB nHL
ABRIS A02 CE-Chirp 40 dB nHL
CE-Chirp
®
stimulus of 40 dB nHL
ABRIS A03 CE-Chirp 45 dB nHL
CE-Chirp
®
stimulus of 45 dB nHL
ABRIS A04 Click 35 dB nHL
Click stimulus of 35 dB nHL
ABRIS A05 Click 30 dB nHL
Click stimulus of 30 dB nHL
ABRIS A06 Click 40 dB nHL
Click stimulus of 40 dB nHL
ABRIS A07 Click 45 dB nHL
Click stimulus of 45 dB nHL
DPOAE D00 2-5 kHz, 3_4, SNR 6 dB
F2 Frequencies: 5k, 4k, 3k, 2kHz (default)
Level (L1/L2): 65/55 dB SPL
F2/F1 Ratio: 1.22
Max. test time: 60 s
# Freq. for pass: 3 / 4
Pass criteria for each frequency:
Min OAE level: -5 dB SPL
Min SNR: 6 dB
Artifact rejection level: 30 dB SPL
DPOAE D01 1.5-4 kHz, 3_4, SNR 6 dB
F2 Frequencies: 4k, 3k, 2k, 1.5k Hz
Level (L1/L2): 65/55 dB SPL
F2/F1 Ratio: 1.22
Max. test time: 60 s
# Freq. for pass: 3 / 4
Pass criteria for each frequency:
Min OAE level: -5 dB SPL
Min SNR: 6 dB
Artifact rejection level: 30 dB SPL
DPOAE D02 1.5-6 kHz, 3_5, SNR 6 dB
F2 Frequencies: 6k, 4k, 3k, 2k, 1.5k Hz
Level (L1/L2): 65/55 dB SPL
F2/F1 Ratio: 1.22
Max. test time: 60 s
# Freq. for pass: 3 / 5
Pass criteria for each frequency:
Min OAE level: -5 dB SPL
Min SNR: 6 dB
Artifact rejection level: 30 dB SPL
DPOAE D05 1.5-6 kHz, 3_6, SNR 7 dB
F2 Frequencies: 6k, 5k, 4k, 3k, 2k, 1.5k Hz
Level (L1/L2): 65/55 dB SPL
F2/F1 Ratio: 1.22
Max. test time: 60 s
# Freq. for pass: 3 / 6
Pass criteria for each frequency:
Min OAE level: -5 dB SPL
Min SNR: 7 dB
Artifact rejection level: 30 dB SPL