Once clear of the screw, pull the
pneumatics forward and rotate to the
left thus providing sufficient space to
remove the cold trap.
Grip the glass flange and remove the
cold trap by gently pulling (a rubber
laboratory glove will provide a better
grip).
DO NOT APPLY EXCESSIVE
FORCE TO THE QUARTZ COLD TRAP
TUBE.
Most difficult with a cold system -
can
switch the unit back on to warm up
before switching off again and
reattempting to withdraw the trap.
Remove the new packed cold trap
from its packaging.
Gently push the trap tube into the cold
trap box.
DO NOT APPLY EXCESSIVE
FORCE TO THE QUARTZ COLD TRAP
TUBE.
If in doubt practice with the trap
alignment tool supplied in the shipping
kit.
NEVER SWITCH UNITY 2 ON
WITH THE TRAP ALIGNMENT
TOOL INSTALLED.