QUI-1057
Page 64
Markes International Ltd
T: +44 (0)1443 230935 F: +44 (0)1443 231531 E: enquiries@markes.com
www.markes.com
1. Place the trap alignment tool into the cold trap
box.
2. The tool should pass smoothly up to the o-ring in
the heated valve before further gentle pressure
moves the tool another 1-2 mm.
If the tool does not move smoothly past the
point where it enters the brass connector
there may be a misalignment problem.
Loosen the 4 screws again and ensure the assembly
is pushed up against the cold trap box.
Reinstall a new cold trap as shown in Section 4.2
1-2 mm
2.
1.