Radical-7 Appendix
www.masimo.com 215 Masimo
Guidance and Manufacturer's Declaration - Electromagnetic
Immunity
Surge
IEC 61000-4-5
+/-1 kV
line(s) to
line(s)
+/- 2 kV
line(s) to
earth
Mains power quality should
be that of a typical
commercial or hospital
environment.
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
100%
dip in
mains
voltage
for 0.5
cycle
60% dip
in mains
voltage
for 5
cycle
30% dip
in mains
voltage
for 25
cycle
Mains power quality should
be that of a typical
commercial or hospital
environment.

Table of Contents