PIPETTOR OVERVIEW
11
Paced Dispense. When you use the paced
dispense feature, the pipettor dispenses the
programmed volume at the specified pace (fast or
slow). Immediately after each dispense step, the
pipettor emits a single beep tone (if beep tone is on)
to prepare you for the next dispense step. See
Section C.5, Operating Procedures.
CAL The calibration of the pipettor is factory set for
distilled water at room temperature. If you are
pipetting fluids with significantly different specific
gravities or temperatures, use this setting to
recalibrate the pipettor. See Section D, Calibration.
SPd
IMPACT
2
only
Once in the special function mode, adjust the
pipetting speed by pressing the Volume key. Five
speed settings are available:
Speed Setting Display
SLOW S
MEDIUM SLOW S
MEDIUM S
MEDIUM FAST S
FAST S F
The current speed setting is always shown at the
bottom of the display. These settings can be selected
and changed at any time before a program step is
performed.