PROGRAMMING AND OPERATION
24
C.4.5 Supernatant Collection
Example: Aspirate 100 µl from 8 rows of a microtiter plate and
discard the total volume.
[Fill]
s
t
Vol
(100) [Enter]
[Fill] [Enter]
[Fill] [Enter]
[Fill] [Enter]
[Fill] [Enter]
[Fill] [Enter]
[Fill] [Enter]
[Fill] [Enter]
[Purge] [Enter] [Enter]
This program is performed with the 850 µl handle. All 100 µl fill
volumes are collected in the same tips. When completed, the total
volume is discarded with the purge step.