USER Manual
33
After pressing the exposure button, a tone warns that X-ray exposure has
begun, and X-ray exposure continues with a corresponding tone. X-rays are
only generated during the period of the consecutive tone and while the button
is held.
While X-rays are generated, the ‘READYsignal on the upper LCD main
control display will change to theEXPOSURE' signal, and the X-ray indicating light
on the front part of the main body will turn yellow.
Once the X-ray exposure has finished, the 'EXPOSURE' signal on the upper LCD
main control display will disappear and the yellow X-ray indicating light will turn
off.
After the X-ray exposure has finished, a ‘WAIT’ signal appears on the
upper LCD main control display and the device will prepare for the next X-ray
exposure over the course of 10 seconds. X
-ray exposure is not available during
the wait period.
CHECK
The modified time set is only saved after an X-ray exposure has occurred.
If, after changing the time set, no exposure is made, the time set reverts
to the default value automatically after the device is turned off.
CAUTION
X-rays are only generated while the X-ray exposure button is being
pressed. If the exposure button is released during the tone, exposure
stops immediately.
CAUTION
Avoid any potential materials which may attenuate the X-ray
beam between the X-ray cone and the target. This could cause image
distortion or a decline in the image quality.

Table of Contents