7.2.3 DX-3000 positioning
The distance between the end of the cone and the patient must be within
at least 20cm (8 inch) from the X-ray focal spot.
The target teeth must be positioned at the center of the cone in order to
cover the whole tooth structure.
Operate the exposure appropriate for the image receptor and the teeth being
radiographed. Ensure the DX-3000’s cone is perpendicular to the receptor and that the
cone is pointed at the center of the teeth.
7.3 X-ray Exposure
7.3.1 X-ray exposure sequence
The DX-3000’s X-ray exposure sequence is as follows. Set up the user and
patient in a location that is optimal for radiography prior to exposure.
1
Set up the exposure time according to the recommendation for the target teeth
or to the user's preference.
2
○ Place the film or digital intra-oral sensor tightly inside the target teeth.
3
○ P
ut the cone of the device up to the target teeth and perpendicular to the film
or intra-oral sensor and
press and hold the x-ray exposure button.
4
Hold the exposure button until exposure has finished and then place
the device
in a safe place.
5
Check images from the fil
m or sensor after shooting.