9 Inspection items before repair request
9.1 Check items before requesting inspection
If an abnormality is found with the product, confirm the following items before requesting
inspection.
Symptom Steps to take
Power defect
After powering on,
tone is not heard and
LCD screen does not
display.
While using (including
X-ray exposure), the
power turns off
automatically.
Confirm that the battery is inserted correctly or
replace it with a charged battery.
Separate the battery from the product to prevent
discharge while not in use.
After checking the remaining battery life, replace it
with a charged battery. If the screen is still not
displayed, contact designated A/S center.
Check
the remaining battery life.
Check the fully charged battery lifetime. If the lifetime
is less than half that of the orig
inal lifetime, replace
the battery with a new battery pack.
The product has an automatic shutoff function to
reduce battery consumption; check it's setting.
Exposure defect
When X-rays are
emitted, but
exposure tone is
not heard.
Check if ‘Remote’ mode is selected at Function Setup
menu. Exposure button operates in ‘Du
al’ mode only.
Check that the LCD display shows main control
screen status. X-rays are emitted with main control
display only.
After powering on,
the tone is heard
but the LCD screen
does not display.