8.3.2 Optimal conditions of use
1.
Temperature: 10 ~ 30
2.
Relative humidity: 30 ~ 60 %
3.
Atmospheric pressure: 900 ~ 1060 Hpa
8.4 Items to check
8.4.1 Frequent inspection items
When turning powering on the product, confirm that the power-on sound is heard
and that the normal initial screen is displayed.
Exposure-time settings are changed by using the time setting buttons. Confirm that
the time may be changed normally.
During X-ray exposure, confirm that the tone is heard, and that the exposure
indicating light turns to yellow.
While charging the battery pack, check the charge indicating light on the cradle to
confirm normal operation.
8.4.2 Periodic inspection items
Confirm that the fully charged battery lifetime is sufficient. Purchase a new
battery pack if the lifetime is half that of original lifetime.
CHECK
The battery pack is a consumable item, and the lifetime is gradually
reduced as it is used. If the lifetime is shortened to less than half of
what it was when it was first used, purchase a new batter pack.