EasyManua.ls Logo

Mecmesin MultiTest 1-dV - Setting the Limit Stops; Test Stand States

Mecmesin MultiTest 1-dV
52 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
22 Mecmesin MultiTest-dV Operating Manual
4.13. Setting the Limit Stops
Limit stops help prevent damage to load cells and fixtures by stopping the crosshead
movement before the moving fixtures come into contact with static parts of the stand.
Their positions are adjusted after the fitting of fixtures and test samples.
There are two manually set limit stops. These are set by loosening the thumbscrew,
moving the stop to a new position and retightening. When the crosshead meets a stop,
it activates a switch. This stops the crosshead movement at an upper or lower limit.
Note: Limit switches are also used to set cyclic test start and end positions (page 37).
4.14. Test Stand States
The test stand can be in one of five states:
1. Test readiness - ready to start, or complete,
2. Testing – test operation sequence is running,
3. Stopped - test interrupted or emergency stop pressed,
4. Jog mode - for jogging or positioning the crosshead manually,
5. Settings menu – for adjusting your test stands settings,
In each state, the selector buttons have functions described by the on-screen icons.
Upper an
d lower limit
stops
Limit stops on a MultiTest
-
dV

Table of Contents

Related product manuals