Mark V ProVis Injection System
15 - 18
ILLEGAL TURRET SIZE SELECTED
The turret selected is not available with the existing injector head con-
figuration. Press RESET, then attempt to select another syringe size.
INCORRECT PASSWORD, TRY AGAIN
An incorrect password was entered in response to the RECALL, 65
procedure of injector head selection. Enter the service access pass
-
word, then select the proper head configuration.
INJECT DELAY
Displayed when the unit is in an injection delay mode prior to actual
fluid delivery.
INJECTING
Injection is in progress.
INJECTION COMPLETE
Injection is finished, the start switch is still closed.
INJECTION NUMBER = XX
A untitled pre-programmed injection has been recalled from memory.
INJECTOR HEAD MALFUNCTION
Actual Flow Rate has exceeded the selected Flow Rate, or, actual
Volume delivered has exceeded the selected Volume.
Procedure:
1. Replace the Power Drive Card.
2. If the condition persists, replace the Servo Control Card.
3. If the condition persists, replace the injector head - suspect
an intermittently open plunger feedback pot.
INTERRUPT TEST FAILED
There is an error on one of the interrupt lines to the CPU. Remove and
reapply power, then attempt to operate the unit again.
Service Procedure:
1. Replace the I/O Card.
2. If the condition persists, replace the CPU Card.
3. If the condition persists, replace the Servo Control Card.
4. If the condition remains, look for an open or grounded
interrupt line on the motherboard.