Troubleshooting Guide
15 - 19
INVALID INJECTION PROGRAM
An attempt was made to store an injection program that has some
invalid parameter(s).
Procedure:
1. Review existing parameters, then correct any errors discov-
ered.
2. Store the injection program.
INVALID PPI NUMBER
Number entered cannot be used. The value is either too large, or a dec-
imal.
I/O: PIO1 INOPERATIVE
Input/Output port 1 (PIO1) on the I/O Card is not functioning properly.
Replace the I/O Card.
I/O: PIO2 INOPERATIVE
Input/Output port 2 (PIO2) on the I/O Card is not functioning properly.
Replace the I/O Card.
LINK ERROR: CAN’T SEND
The Control Panel if it is unable to transmit to the Main Unit. Remove
and reapply power.
Service Procedure:
1. Check the Remote Panel cable for broken or intermittent
connections.
2. If the condition continues, replace the PPI card in the Main
Unit.
3. If the condition persists, call MEDRAD or the local dealer for
assistance.
MAX DELAY IS 99.9 SECONDS
A delay above 99.9 was entered. Press RESET, then enter a value no
greater than 99.9 seconds.
MAX FLOW IS 59 ML/MIN OR ML/HR
Injector is in the ML/MIN or ML/HR Flow scale, and a Flow Rate above
59 ml/min or ml/hr was entered. Press RESET, then enter a value no
greater than 59 ml/min or ml/hr.
MAXIMUM DURATION IS 650 SECONDS
A Duration above 650 seconds was entered. Press RESET, then enter a
Duration value no greater than 650 seconds.