EasyManua.ls Logo

MELAG DAC Premium Plus - Wastewater Connection Methods; Initial Test Runs and User Training

MELAG DAC Premium Plus
32 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3 Setup and installation
17
Connection to the wastewater
Method 1: Via the one-way drain
If you connect the one-way overflow hose of the device to the siphon of the domestic water supply, you need a drain
hose with a length of 2m (art. no.ME36585, not included in the scope of delivery).
Method 2: Via the pressure release
To connect the device directly to the wastewater, i.e. from the pressure outlet directly into the siphon of the domestic
water supply, you need a drain hose with a length of 1m (art. no. ME39180, not included in the scope of delivery).
Setting up the steam sterilizer
For a fault-free operation, the steam sterilizer must be set up horizontally by using a water level located close to the
sterilization-chamber flange. Then the front feet of the unit must be unscrewed by about five rotations in order to give the
steam sterilizer a slight backwards tilt.
Test runs
Vacuum test with cold sterilization chamber
Perform aVacuum test with an empty cold sterilization chamber and record the result.
Universal-Program
If the vacuum test was successful, run aUniversal-Program with 1.5kg load (instruments) and record the result.
Instructing the users
Explain all the user-typical features for the documentation and setting combinations for the operator.
Hand over the manufacturer’s inspection report. The declaration of conformity of the Pressure Equipment Directive and
the Medical Device Directive are included in the manufacturer's inspection report.

Related product manuals