Putting into Operation
23 D39303 02/2024
■ Conductivity of approx. >1; <10µS/cm
■ neutral pH value (between 5 and 7)
■ Chlorine-free
The use of double-distilled water / ultrapure water / other highly purified water (a variety of
terms are also commonly used) with an electrical conductance below about < 1µS/cm
must be avoided. The use of such water is not necessary and could damage the unit by
corroding metallic components on and in the unit. Unsuitable water with an electrical
conductance greater than 10µS/cm will damage the unit due to the residues that occur
during evaporation and vaporisation, including the formation of limescale deposits.
5.4 Connecting and Filling the Water Tank
1. Fill the water tank with water.
2. Connect the enclosed tube to the “H
2
O” connection on the rear of the cabinet.
5.5 Switching on Unit
1. Switch on the appliance by pressing the main switch on the front of the appliance.
ð The starting process is shown by three animated white dots (see
}7.1Warning Message of the Monitoring Function).
i
After the first start-up, the appliance display is set to English by default. You can change
the language as described in chapter }8.2 Basic Operation in Menu Mode Using the
Example of Language Selection. However, to get a basic overview of operating the
appliance, you should read the following chapter first.
See also
2 Basic Operation in Menu Mode Using the Example of Language Selection [}42]
2 Warning Message of the Monitoring Function [}38]