Operation and Control
35 D39303 02/2024
9. Press the confirmation key to confirm.
ð Temperature monitoring is now active.
i
In menu mode you can set, whether an alarm should be accompanied by an acoustic
signal (see }8.7Acoustic Signals).
See also
2 Acoustic Signals [}51]
6.6.7 Humidity Monitoring
If humidity monitoring has been triggered, this is indicated by the humidity display: the
actual humidity is highlighted in red and is accompanied by a warning symbol .
If the acoustic signal has been activated in Menu mode to accompany an alarm ( }
8.7 Acoustic Signals, which is indicated by the speaker symbol ), the alarm is
accompanied by an intermittent acoustic signal. Information on what to do in this case is
provided in the chapter }7Malfunctions, Warning and Error Messages.
Adjusting humidity monitoring
1. Press the activation key to the left of the ALARM display.
ð The temperature monitoring setting is automatically activated.
2. Turn the turn control until the humidity setting is highlighted.
3. Confirm the selection by pressing the confirmation key.
ð The lower humidity alarm limit is automatically highlighted.
4. By turning the turn control, set the required lower alarm limit, 50% rh in the example on
the left.
5. Confirm the selection by pressing the confirmation key.
ð The upper humidity alarm limit is automatically highlighted.
6. By turning the turn control, set the required upper alarm limit, 70 % rh in the example
on the left.