MI 3295 Step Contact Voltage Measuring System Technical specifications
50
Charger socket input voltage .................. 12 V (10 %)
Charger socket input current .................. 400 mA max
Battery charging current ......................... 250 mA (internally regulated)
Measuring category ................................ CAT IV / 50 V
Protection classification .......................... double insulation
Pollution degree ..................................... 2
Degree of protection ............................... IP 40
Display .................................................. 128 x 64 dots matrix display with backlight
Memory .................................................. 1500 memory locations
Communication interface ........................ RS232, USB
Dimensions (w×h×d)............................... 23 cm × 10.3 cm × 11.5 cm
Weight .................................................. 1.3 kg (with batteries)
Environmental conditions
Reference temperature range ................. 10 C … 30 C
Reference humidity range ...................... 35 % … 65 % RH
Operation conditions
Working temperature range .................... 0 C … +40 C
Maximum relative humidity ..................... 85 % RH (0 C … 40 C), non-condensing
Storage conditions
Temperature range ................................. -10 C … +60 C
Maximum relative humidity ..................... 90 % RH (-10 C … +40 C)
80 % RH (40 C … 60 C)
Accuracies apply for 1 year in reference conditions. Temperature coefficient outside these
limits is 0,2 % of measured value per C, and 1 digit.