EasyManuals Logo

METREL TeraOhm 5kV Plus User Manual

METREL TeraOhm 5kV Plus
48 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #47 background imageLoading...
Page #47 background image
MI 3201 TeraOhm 5 kV Plus Specifications
47
Display range Test voltage (V) Resolution Accuracy
0 5500 V
1 V
(3 % of reading + 3 V)
Withstanding voltage DC
DC test voltage:
Voltage value: 250V to 5kV, steps 25V.
Accuracy: -0 / +10 % + 20 V.
Display range Test voltage (V) Resolution Accuracy
0 5500 V
1 V
(3 % of reading + 3 V)
Leakage current
Display range Itrigg (mA) Resolution Accuracy
0.000 0.009 1 A (3 % of reading + 3 digits)
0.01 5.50 10 A (3 % of reading)
Voltage
Voltage AC or DC
Display range External Voltage (V) Resolution Accuracy
0 600
1 V
(3 % of reading + 4 V)
Frequency of external voltage
Display range (Hz) Resolution Accuracy
0 and 45 65
0.1 Hz
0.2 Hz
Note:
- for frequency between 0 and 45 Hz displayed <45 Hz
- for frequency over 65 Hz displayed >65 Hz
- for voltages under 10V frequency result is shown as ---
Input resistance: 3 M 10 %
Capacitance
Measuring range C: 50 F*
Display range C Resolution Accuracy
0.0 99.9 nF
0.1 nF
100 999 nF
1 nF
1.00 50.00 F
10 nF
(5 % of reading + 4 nF)
C measured if R > 5 M
*Full-scale value of capacitance is defined according to the following equation:
C
FS
= 10 F * U
test
[kV]

Questions and Answers: