Technical characteristics
48 100,000 cts digital multimeters
Technical characteristics of the MTX 3292, MTX 3293 (continued)
With this function, the user can measure the true RMS (TRMS) value of an AC voltage with its
DC component (no capacitive coupling) or without its DC component.
The 100mV range is available only in Manual mode, by " ".
In the VAC & VAC+DC modes and for signals >1kHz, the range of uncertainty displayed is
given for information only: we recommend using the formulas below.
VLowZ: The error should be slightly greater than the error in VAC.
() not contractual indicative values (see curves below)
() limitation at high frequency
() BP: Freq [kHz] limited to: 15,000/U input [V]
U input [V] limited to: 15,000/Freq [kHz]
Example: U input = 1,000 VAC Max. frequency: 15,000/1,000 = 15kHz
In the presence of a DC component: additional error: (UDC/U measured)x(0.7% + 70 D)
Example: UDC = 2V, U measured = 5 Vrms Additional error: 0.28% + 28 D
- Rejection: common mode >80dB at 50Hz or 60Hz depending on selection
- Automatic or manual selection of the ranges
- Protection by varistors
- Maximum acceptable permanent voltage: 1,414 Vpk
- Specifications valid from: 10 to 100% of the range in the band from 20kHz to 100kHz
- Influence of the peak factor on the accuracy in VAC, VAC+DC at 50% of the range:
1% for a peak factor <3.
As soon as the PEAK symbol appears, use the AUTO PEAK mode.