EasyManua.ls Logo

Metrix MTX 3292 - Surveillance Mode

Metrix MTX 3292
63 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
How are the various quantities measured?
42 100,000 cts digital multimeters
How are the various quantities measured? (continued)
11. Surveillance
mode
The SURV mode (available under MEAS…)
monitors the variations of a signal, recor-
ding the extremes (MIN and MAX) of the
main measurement and calculating its
mean (AVG).
For each quantity stored, the multimeter
records the corresponding date and time.
This mode is active for the following
functions:
V, Hz, Ohm, clamp, capacitance,
temperature, and current.
Integration time at
least 200ms,
programmable
according to your
configuration: Start
Stop, then look-
up of the quantities
on the screen, in a
specific window.
It is possible to recover a screen grab of
this window under our SX-DMM software,
but this mode cannot be stored in the
instrument.
Reset of the MIN/MAX values by a
long press on MEAS…
PEAK
The rapid peak measurements are
available in the MEAS, MEAS+, PK+ and
PK- secondary measurements for the
following measurement functions:
V and A (AC, AC+DC); integration time
less than 250µs.
Reset of the values by a long press
on MEAS…