Appendix
62 100,000 cts digital multimeters
APPENDIX
Default
configuration
In User mode, the device restarts in the user's personal configuration (General and
Measurement menus) and the function selected when switched off, but coupling in
Volt function (AC+DC).
In Basic mode, the default, the multimeter starts up in its elementary configuration
(default values) and in the Volt function (AC+DC).
General
Language: yes
Beep: yes
Sleep: yes
Lighting: ECO Communication: IR
IR baud: 38400 Configuration: basic
Energy: Ni-MH
Storage battery capacity: 2,400mAh
Communication protocol: MODBUS
Measurement
Filter: NO Impedance: 10/20M
dBm REF: 600 W REF: 50
Func. CLAMP,
Func. MATH
Function: V Unit: A
Ratio: 1A/AV
Function: V Unit: none
Coef. A: 1 Coef. B: 0
Func. MEM
Recording interval: 1s
Nb. of records 3292: 1000 Nb. Of records. 3293: 6500
Main functions
V, A: AUTO, AC+DC Hz: 10V range
, Capacity: AUTO ° C: ° C, Pt 100
Restart configuration assumes no leads connected. If they are connected,
they will be taken into account in the selection of the function.
Instructions before
recharging the
storage batteries
Before recharging, check that the device is equipped with all 4 storage batteries.
It is not necessary to withdraw them to recharge them. If "Ni-MH" is selected in the Type
of Energy menu (see paragraph), then charging is enabled.
An attempt to charge with primary batteries in the device may damage the device.
For safety reasons, the charging of the storage batteries is enabled only between:
0°C and 35°C.
Note: an elevation of the internal temperature by a current measurement may possibly trip
the thermal safety.
In order to keep the storage batteries in good condition, use the multimeter until the min.
level is reached before recharging.
Then connect the plug of the power supply unit (12 VDC) to the specific jack (illustration of
the front panel).
Connect the power supply unit (12 VDC) to mains.
The symbol opposite on the display unit is used to monitor the course of the charging with
a % of charge:
- battery charged green symbol and 100%
- battery discharged orange symbol, indicating charging recommended
- battery level at limit red symbol and xx%
- battery level too low blinking red symbol and % together with audible beep
The storage batteries are fully charged when the symbol is stabilized with 4 segments
(each fixed plateau is acquired), after approximately 6 h.
The multimeters are delivered with 2400mAh Ni-MH storage batteries.
These used storage batteries must be turned over to a recycling company or a company
processing hazardous materials.
Never throw these storage batteries away with other solid waste.
For more information, contact your Manumesure agency.
Once the storage batteries are fully charged, the device stops automatically.
When the multimeter is delivered, it may happen that the storage batteries are discharged
and require full recharging.