How are the various quantities measured?
32 100,000 cts digital multimeters
How are the various quantities measured? (continued)
3. Current
measurement
with clamp
Key:
To avoid opening a circuit, we recommend measuring the current with a current
clamp, output A or V (Ax function).
To do this, proceed as follows:
a) Activate the clamp function and, depending on the type of clamp connected,
double-press "Clamp" or use the setup/clamp menu.
b) Select the type of Measurement clamp output (V, A)
c) Define ratio A displayed on the clamp Val1/Val2 or xxxx.Xa/xxxx.Xv
(by default 1A/1V) to be incorporated (Ok to Validate, or Cancel).
d) Define the physical unit to be displayed (default: A): 3 programmable fields
Reminder
The clamp function includes a precise ratio xxxx.XA/xxxx.XV or XA, making it
possible to connect a broad range of current clamps that you will find in the
CHAUVIN ARNOUX catalogue; it is however necessary to verify that the
input/output range of the clamp matches the ranges available on the multimeter.
The accuracy of this "clamp" function depends on the accuracy of the clamp and
of the range used on the multimeter.
Connections
possible in
current
measurement:
a)
Current clamp
with current output
connected to the
multimeter
b)
Current clamp
with voltage output
connected to the
multimeter
or:
Measurement in
series in a circuit
No secondary
measurements
in this quantity