6.3 User methods
794 Basic Titrino
181
- Excess of dispensed titrant is determined..
- Transfer of the determined excess of titrant to
P2O5Fert.
P2O5-2 Titration of dihydrogenphosphate.
- Transfer of the determined end point as tem-
porary variable to P2O5Fert.
'pa
794 Titrino 01102 794.0010
date 2002-01-03 time 10:04 0
DET pH P2O5-2
parameters
>titration parameters
meas.pt.density 4
min.incr. 10.0 µl
titr.rate max. ml/min
signal drift 50 mV/min
equilibr.time 26 s
start V: OFF
pause 0 s
meas.input: 1
temperature 25.0 °C
>stop conditions
stop V: abs.
stop V 99.99 ml
stop pH OFF
stop EP 1
filling rate max. ml/min
>statistics
status: OFF
>evaluation
EPC 5
EP recognition: window
low lim.1 pH 7
up lim.1 pH 9
low lim.2 pH OFF
fix EP1 at pH OFF
pK/HNP: OFF
>preselections
req.ident: OFF
req.smpl size: OFF
activate pulse: OFF
------------
'de
794 Titrino 01102 794.0010
date 2002-01-03 time 10:04
DET pH P2O5-2
def
>formula
>silo calculations
match id: OFF
>common variables
>report
>mean
MN1=RS1
>temporary variables
C71=EP1
------------
'de
794 Titrino 01102 794.0010
date 2002-01-03 time 10:03
DET pH P2O5-1
def
>formula
excess=C41-EP1
RS1 text excess
RS1 decimal places 3
RS1 unit: ml
>silo calculations
match id: OFF
>common variables
>report
>mean
MN1=RS1
>temporary variables
C70=RS1
------------

Table of Contents